SUMMARY
In this work, ELS-Detector has been used for the development of an HPLC method for the determination of clarithromycin.  Isocratic reversed phase HPLC approach has been developed using a C-18 column (Waters Spherisorb 5 μm ODS2, 4.6x250 mm) and a mobile phase consisting of acetonitrile / aqueous trifluoroacetic acid as pairing reagent. Experimental parameters (temperature of heated drift tube, flow rate of mobile phase, gas flow rate, mobile phase composition) were optimized. Clarithromycin’ s stability was thoroughly examined in different solvent systems. Using the optimized conditions the working range was 5-100 μg/mL (upper limit can be increased considerably), with a detection limit of 4.5 μg/mL (6x10-6 M). The method was validated as per ICH guidelines. The retention time is 4.7 min. The method was successfully applied for the content assay of clarithromycin formulations.Â